Combined APPI/APCI Source
2-3
Switching on the UV lamp (from the MassLynx™ Tune window) provides a
constant photon output. The intensity of incident radiation upon the sample
molecules is varied by manually adjusting the distance between the UV lamp
and probe tip, using the handle on the APPI lamp drive assembly to move the
whole assembly on its mounting shaft. The mounting shaft’s indexing
numbers aid in setting this distance.
See also:
“To obtain an ion beam and tune the instrument for APPI mode” on
.
APPI lamp drive assembly and side plate
Indexing numbers
HV electrical connection
APPI electrical connection
APPI lamp drive assembly
Handle
Thumbscrew (×3)
Source enclosure
door securing clip
Captive securing
screw (×4)
Mounting shaft
APPI source enclosure
side plate