VTI Instruments Corp.
294
EX10SC Signal Conditioning
EX10SC-8B34-04
S
PECIFICATIONS
D
ESCRIPTION
2- and 3-wire 100 Ω RTD (0 °C to 600 °C)
I
NPUT
R
ANGE
L
IMITS
Input Range
Accuracy
0 °C t o +600 °C (+32 °F to +1112 °F)
±0.45 °C
I
NPUT
R
ESISTANCE
Normal
Power Off
Overload
50 MΩ
200 kΩ
200 kΩ
I
NPUT
P
ROTECTION
Continuous
Transient
240 V ac
ANSI/IEEE C37.90.1
S
ENSOR
E
XCITATION
C
URRENT
0.25 mA
L
EAD
R
ESISTANCE
E
FFECT
±0.02 °C/Ω
CMV
(C
OMMON
M
ODE
V
OLTAGE
)
Input to Output
1500 V rms maximum
T
RANSIENT
Input to Output
ANSI/IEEE C37.90.1
CMR
(C
OMMON
M
ODE
R
EJECTION
)
50 Hz or 60Hz
120 dB
NMR
(N
ORMAL
M
ODE
R
EJECTION
)
70 dB at 60 Hz
A
CCURACY
See ordering information
S
TABILITY
Offset
Gain
±20 ppm/°C
±50 ppm/°C
N
OISE
Output, 100 kHz
200 μV rms
B
ANDWIDTH
(-3 dB)
3 Hz
R
ESPONSE
T
IME
90% span
150 ms
RTD
S
TANDARDS
100 Ω Pt
Alpha Coefficient
DIN
JIS
IEC
0.00385
DIN 43760
JIS C 1604-1989
IEC 751