VTI Instruments Corp.
108
EX10xxA Programming
an array of four 8-bit values representing the enabling of events from any of the 8 channels of
the digital I/O port. The order of the values is: positive edge, negative edge, positive level,
negative level. Each value is specified in either decimal (0 through 255) or hex (0x00 through
0xFF). Within the 8-bit field, the MSB corresponds to DIO channel 7 and the LSB
corresponds to DIO channel 0.
an array of four 8-bit values representing the enabling of events from any of the eight LAN
events. The order of the values is: positive edge, negative edge, positive level, negative level.
Each value is specified in either decimal (0 through 255) or hex (0x00 through 0xFF). Within
the 8-bit field, the MSB corresponds to LAN 7 and the LSB corresponds to LAN 0.
a Boolean value indicating the enable status of the Timer event
a Boolean value indicating the enable status of the Immediate event.
a Boolean value indicating the enable status of the LXI alarm event
Regardless of this setting, software arms are always enabled.
Example #1: This code block enables arm on the timer only.
ViUInt16 vtb_masks[4] = {0,0,0,0};
ViUInt16 dio_masks[4] = {0,0,0,0};
ViUInt16 lan_masks[4] = {0,0,0,0};
vtex10xxA_set_arm_sourceEx(vi, vtb_masks, dio_masks, lan_masks, 1, 0, 0);
Example #2: This code block enables arm on a positive level on DIO channels 0 through 3 and a
negative edge on VTB channel 6.
ViUInt16 vtb_masks[4] = {0,64,0,0};
ViUInt16 dio_masks[4] = {0,0,0x0F,0};
ViUInt16 lan_masks[4] = {0,0,0,0};
vtex10xxA_set_arm_sourceEx(vi, vtb_masks, dio_masks, lan_masks, 0, 0, 0);
Example #3: This code block enables software arm only.
ViUInt16 vtb_masks[4] = {0,0,0,0};
ViUInt16 dio_masks[4] = {0,0,0,0};
ViUInt16lan_masks[4] = {0,0,0,0};
vtex10xxA_set_arm_sourceEx(vi, vtb_masks, dio_masks, lan_masks, 0, 0, 0);
The current ARM source is queried with the
query.
The ARM count is configured with the
function, specified with a value
from 1 to (2
31
-1). This value is reset with each trigger initialize or automatically upon reaching zero
when init continuous is enabled.
Example: This code block sets an ARM count of 10.
vtex10xxA_set_arm_count(vi, 10);
The current ARM count is queried with the
query.
Alternatively, the ARM count can be set to infinity, overriding any manual setting of ARM count,
with the
function. This function simply accepts a Boolean value
indicating the enable status of an infinite ARM count.
The current setting of an infinite ARM count is queried with the
query.
Summary of Contents for EX1000A
Page 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Page 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Page 34: ......
Page 56: ......
Page 74: ......
Page 130: ......
Page 282: ......