www.vtiinstruments.com
EX10xxA Web Page Operation
89
If a PTP master clock does not exist on the network, an LXI device on the network will begin serving
as the PTP time master. The
Time Configuration
page allows for PTP time to be set manually.
From the
Time Configuration
page, select the
Set Time
checkbox, then select the
Grandmaster
radio button from the
Target
field, and then configure the time by hand. Note that the
Time
field
should be completed using a 24-hour format.
To use the EX10xxA as the PTP clock that will serve time on the LXI network, select the
Self
radio
button in the
Target
field. The EX10xxA then becomes the PTP master clock. While this ensures
measurement synchronicity between the units that to this clock, the overall timestamp accuracy is
not as accurate because of the time’s hand-set nature. For higher overall timestamp accuracy, use a
PTP master clock that is synchronized to a global time source like, such as GPS.
Using SNTP/NTP Time
F
IGURE
5-7:
T
IME
C
ONFIGURATION
P
AGE
-
SNTP
By selecting
SNTP
from the
Time Source
menu, the EX10xxA can be synchronized with an SNTP
server. The
Time
Zone
drop menu allows the user to select the local time zone. SNTP, or NTP
(Network Time Protocol), servers are specified in the
NTP Servers
field, by IP address or hostname.
Once configured, the user must click
Submit
in order for the changes to take effect.
To configure the EX10xxA to use NTP time as its clock source:
1)
Select
SNTP
from the
Time Source
field. The page will change so that it can be configured.
2)
Add
/
Remove
NTP servers use the appropriate buttons until the EX10xxA is configured to
communicate to the desired server(s).
3)
Click the
Save
button to save the server configuration.
4)
Click
Submit
to begin using NTP time.
NOTES
1) If the
Submit
button is clicked before saving an NTP configuration, the EX10xxA will start
using NTP time, but will use the last previously saved server configuration.
2) Specifying SNTP or NTP servers by hostname requires that the instrument has DNS, either by
DHCP or Static IP.