www.vtiinstruments.com
EX10xxA Calibration
305
F
IGURE
10-2:
EX1044
C
ABLING
D
IAGRAM
NOTE
If the
Agilent 3458A is used, the “Auto Cal All” procedure should be performed prior to calibrating
the EX10xxA.
C
ONNECTING TO THE
EX10
XX
A/RX10
XX
/EX1044
There are three ways to connect to the EX10xxA/RX10xx: using the IP address of the
EX10xxA/RX10xx, using Bonjour for auto-discovery, or using the Agilent Connection Expert
(ACE). For more information on obtaining the EX10xxA/RX10xx IP address and using this as the
connection method, refer to the
. To connect to the
EX10xxA/RX10xx via Bonjour, refer to
To use ACE, use the following steps:
1)
Open ACE and click on the
Add instrument
toolbar.
2)
Select
Add LAN instrument on LAN
and then click
OK
. ACE will search the network and
list the LXI instruments connected to the LAN.
F
IGURE
10-2:
A
DD
I
NSTRUMENT
I
NTERFACE
3)
To simply view the EX10xxA/RX10xx web page, click on its webpage button. If desired, the
EX10xxA/RX10xx can be added to ACE as a LAN instrument by enabling the appropriate
Select
checkbox and then clicking
OK
.
Summary of Contents for EX1000A