VTI Instruments Corp.
158
EX10xxA/RX1032/EX1044 Function Calls
vtex10xxA_get_channel_range
F
UNCTION
P
ROTOTYPE
ViStatus _VI_FUNC vtex10xxA_get_channel_range(ViSession
vi
, ViInt32
channel
, ViPReal64
range
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
channel
= the channel for which the range value is desired. Valid input values: 0 through 47.
range
= returns the voltage range value for the specified
channel
. Valid return values: 0.01 (10 mV), 0.067
(±67 mV), 0.1 (100 mV), 1 (1 V), and 10 (10 V).
D
ATA
I
TEM
R
ESET
V
ALUE
Not applicable to this function.
D
ESCRIPTION
This function returns channel gain of a voltage channel. An error is returned if a thermocouple channel is specified
by the
channel
parameter.
E
XAMPLE
Summary of Contents for EX1000A
Page 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Page 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Page 34: ......
Page 56: ......
Page 74: ......
Page 130: ......
Page 282: ......