www.vtiinstruments.com
EX10SC Signal Conditioning
295
EX10SC-8B36-04
S
PECIFICATIONS
D
ESCRIPTION
Potentiometer input (0 Ω to 10 kΩ)
I
NPUT
R
ANGE
0 to 10 kΩ
I
NPUT
R
ESISTANCE
Normal
Power off
Overload
50 MΩ
200 kΩ
200 kΩ
I
NPUT
P
ROTECTION
Continuous
Transient
240 V ac
ANSI/IEEE C37.90.1
S
ENSOR
E
XCITATION
C
URRENT
100 Ω, 500 Ω, 1 kΩ sensor
10 kΩ sensor
0.25 mA
0.10 mA
L
EAD
R
ESISTANCE
E
FFECT
100 Ω, 500 Ω, 1 kΩ sensor
10 kΩ sensor
±0.01 Ω/Ω
±0.02 Ω/Ω
CMV
(C
OMMON
M
ODE
V
OLTAGE
)
Input to output
1500 V rms maximum
T
RANSIENT
Input to output
ANSI/IEEE C37.90.1
CMR
(C
OMMON
M
ODE
R
EJECTION
)
50 Hz or 60Hz
120 dB
NMR
(N
ORMAL
M
ODE
R
EJECTION
)
70 dB at 60 Hz
A
CCURACY
±0.05% span
L
INEARITY
±0.02% span
S
TABILITY
Offset
Gain
±20 ppm/°C
±50ppm/°C
N
OISE
Output, 100 kHz
200 μV rms
B
ANDWIDTH
(-3 dB)
3 Hz
R
ESPONSE
T
IME
90% span
150 ms