88
DSGH Installation and Operation Guide
Diagnostics and repair
Procedure 5.2: To change the due date of source wipe or
shutter check
Field repair procedures
Very few parts are field repairable, but you can replace entire assemblies or
boards. These parts are replaceable:
• CPU circuit board
• Power supply circuit board
Spare parts
Contact VEGA Field Service at
+1 513-272-0131
for parts, service, and
repairs.
Outside the U.S., contact your local VEGA representative for parts, service,
and repairs.
Replacing the CPU or power supply board
You may have to replace a circuit board if there is damage to one of its
components. Before replacing a circuit board, check the troubleshooting
flowcharts or call VEGA Field Service to be sure a replacement is
necessary.
The sensor EEPROM contains a backup of the CPU board EEPROM. After
replacing the CPU board, you must perform a memory backup to update the
CPU board’s EEPROM with the information in the sensor board EEPROM.
1.
Select
Setup | Source functions
.
2.
Change the number of days in the
Wipe Interval
or
Shutter
Check Interval
field.
3.
Click
OK
.
Note: Use great care to prevent damage to the electrical components
of the gauge. VEGA recommends appropriate electrostatic
discharge procedures.
Summary of Contents for DSGH
Page 2: ......
Page 4: ...2 DSGH Installation and Operation Guide NOTES ...
Page 8: ...6 DSGH Installation and Operation Guide NOTES ...
Page 12: ...10 DSGH Installation and Operation Guide Preface NOTES ...
Page 32: ...30 DSGH Installation and Operation Guide Installation Mounting the DSGH gauge ...
Page 56: ...54 DSGH Installation and Operation Guide Calibration NOTES ...
Page 94: ...92 DSGH Installation and Operation Guide Diagnostics and repair NOTES ...
Page 98: ...96 DSGH Installation and Operation Guide NOTES ...
Page 99: ......