DSGH Installation and Operation Guide
Advanced functions
Table 4.1
Process Chain tab — display values
Value
Description
Sensor Temperature
The internal probe's measurement of the sensor
temperature.
Sensor Counts
True counts output from the sensor, but before
application of:
• Temperature compensation
• Standardize
• Sensor uniformity gains
Temp Comp Counts
The temperature-compensated counts that are sensor
counts with application of temperature compensation.
Raw Counts
Temperature compensated counts with application of
uniformity gain.
Adjusted Counts
Sum counts that are raw counts plus auxiliary raw
counts. In most applications, this does not use auxiliary
input, so sum counts = raw counts.
Source Decay Counts
Sum counts with application of source decay gain.
Standardize Counts
Displays standardize counts that are source decay
counts with application of standardization gain.