Calibration
During the standardization procedure, the system displays:
• A default value for the standardization condition
• A prompt to enter the actual process value of the standardization
condition
Standardization reminder
If you enable the standardization due alarm, the gauge produces an alarm
when standardization is due. The standardize interval is programmed in
Setup | Cal Parameters
.
You can perform a standardization using any of the following:
• Process
• Water (or other repeatable fluid)
• Absorber plates
Standardization on water
Standardization on water is the best choice if water is readily obtainable.
For example, if the process is water based or if the process is slurry with
water as the carrier.
Advantages: If the source of water is of consistent purity (for example, city
vs. pond), standardization on water is highly accurate. Water is a good
choice if the density of the process is around 1.0SpG because
standardization on water would be within the process span. This ensures
high accuracy at the point on the calibration curve.
Disadvantages: You must be able to empty and fill the process pipe with a
consistent source of water.
Note:
This method requires you to select Default Value and enter the
Standardize Default Value on the SetUp | Cal Parameters tab.
Summary of Contents for DSGH
Page 2: ......
Page 4: ...2 DSGH Installation and Operation Guide NOTES ...
Page 8: ...6 DSGH Installation and Operation Guide NOTES ...
Page 12: ...10 DSGH Installation and Operation Guide Preface NOTES ...
Page 32: ...30 DSGH Installation and Operation Guide Installation Mounting the DSGH gauge ...
Page 56: ...54 DSGH Installation and Operation Guide Calibration NOTES ...
Page 94: ...92 DSGH Installation and Operation Guide Diagnostics and repair NOTES ...
Page 98: ...96 DSGH Installation and Operation Guide NOTES ...
Page 99: ......