DSGH Installation and Operation Guide
53
Calibration
Procedure 3.7: To standardize the gauge
Note:
You cannot use absorber plates for the initial process calibration.
You must determine the equivalent value of the absorber plates
after the initial process calibration of the density gauge when the
gauge was mounted to the pipe.
1.
Select
Calibration | Standardize
.
2.
Click
Execute
.
3.
Click
OK
.
4.
Enter the reading.
5.
Click
Start
.
6.
Click
Accept
.
7.
Enter the process value.
8.
Click
OK
.
9.
Click
OK
.