DSGH Installation and Operation Guide
29
Installation
Avoid source cross-talk
When multiple adjacent pipes or vessels have nuclear gauges, you must
consider the orientation of the source beams so each gauge senses
radiation only from its appropriate source.
The best orientation, in this case, is for the source holders to be on the
inside with radiation beams pointing away from each other.
Mounting the measuring assembly
You can mount the density gauge on the pipe by positioning the detector
housing and source holder brackets with the bolts that VEGA provides. If
the pipe has insulation, the density gauge and source holder should
have external support to prevent crushing of the insulation.
Note: In some cases, the handle on the source holder operates a
rotating shutter. When installing or removing the assembly from
the pipe, you must turn the handle to the closed (OFF) position
and lock the handle with the combination lock provided.