Pattern
Independent TX/RX test Pattern selection. PRBS 2^23-1 (normal or inverted) or PRBS 2^31-1 (normal or inverted)
Measurements
Measurements Setup
Configure a test to run for a fixed duration or a delayed start.
Mode
: Manual, Timed, and Auto selections are available.
Manual:
This is linked to the Start/Stop function.
Timed:
The test duration can be set by the user. The test duration can be set in seconds, minutes, hours or days. The test
is activated by the Start/Stop function.
Auto:
A predetermined start time can be set by the user. The test duration can be set in seconds, minutes, hours or days.
After programming the start time and duration, press the Startbutton and the test will be activated automatically when the
programmed start time is reached.
Interval Save/Results Auto Save and Auto Server Upload is available in all modes. Depending on the V300 unit and test module
installed, Interval Save will appear as Results Auto Save.
Interval Save or Results Auto Save:
Test results automatically save at a specified interval. To enable interval save,
select
ON
and enter the
Save Interval
(in minutes).
Auto Server Upload
: Automatically uploads results to the server. The unit needs to be connected and registered to an R-Server
or FTP-Server to use this function. Selecting a server type will display a message prompt with the option to connect and register
to the server.
RXT-3900 e-Manual D07-00-111P Rev A00
Summary of Contents for RXT3900
Page 1: ...RXT 3900 e Manual D07 00 111P Rev A00...
Page 19: ...Go back to top Go back to TOC RXT 3900 e Manual D07 00 111P Rev A00...
Page 50: ...RXT 3900 e Manual D07 00 111P Rev A00...
Page 74: ...RXT 3900 e Manual D07 00 111P Rev A00...
Page 75: ...RXT 3900 e Manual D07 00 111P Rev A00...
Page 119: ...Go back to top Go back to TOC RXT 3900 e Manual D07 00 111P Rev A00...
Page 240: ...Go back to top Go back to TOC RXT 3900 e Manual D07 00 111P Rev A00...
Page 282: ...Go back to top Go back to TOC RXT 3900 e Manual D07 00 111P Rev A00...
Page 294: ...Go back to top Go back to TOC 13 15 IP Monitoring RXT 3900 e Manual D07 00 111P Rev A00...
Page 296: ...Net Wiz Trace Route Setup Go back to top Go back to TOC RXT 3900 e Manual D07 00 111P Rev A00...
Page 363: ...Go back to top Go back to TOC RXT 3900 e Manual D07 00 111P Rev A00...
Page 377: ...Go back to top Go back to TOC RXT 3900 e Manual D07 00 111P Rev A00...