Ethernet
© VeEX Inc. All Rights Reserved.
174
Test Module. This port is marked CLK on the connector panel.
Avoid using rigid BNC-to-SMA adapters to prevent any stress on the test set’s connector. Flex-
ible adapters or cables are recommended.
l
Recovered Clock Output
: The reference clock used by the SyncE master or slave can be regen-
erated out of the PDH TX port (marked Tx on the connector panel) with a different clock format in
order to synchronize other network elements. In Slave mode the Reference Clock Output is the
regenerated clock recovered by the SyncE slave.
The clock can be formatted to: 2Mbps (E1 signal), 2MHz, 10MHz, 25MHz, 125MHz, 1PPS and
None.
l
If 2Mbps clock is selected from Recovered Clock Output, then the following parameters need to be
set:
l
Line Code:
HDB3 or AMI
l
Framing:
Unframed, PCM31, PCM31C, PCM30, or PCM30C
l
PRBS Pattern
l
Invert
l
Offset(ns)
(only for 1 PPS Recovered Clock Output)
Press
Apply
once all the parameters are set.
Status
Summary of Contents for MTX150 Series
Page 1: ...MTX150x User Manual Multi Service Installation Maintenance Test Set P N D07 00 148P Rev B00 ...
Page 11: ...Customer Care 303 Table of Contents VeEX Inc All Rights Reserved 11 ...
Page 171: ...171 VeEX Inc All Rights Reserved Ethernet V FTP Setup Download V FTP Download Results SyncE ...
Page 246: ...Common Functions VeEX Inc All Rights Reserved 246 Bluetooth Devices Bluetooth Scan ...
Page 248: ...Common Functions VeEX Inc All Rights Reserved 248 Bluetooth Devices Bluetooth Scan ...
Page 267: ...267 VeEX Inc All Rights Reserved Common Functions ...