5. Installing the Equipment: Wall Mount
78
Green 16/18 Installation Manual
5.3.4 Drilling Holes on the Wall
1. Drill the wall holes of size 10.5 mm x 30 mm (depth) using the concrete hammer
drill.
2. Remove the debris and clean the holes using the dust pump.
3. Using the hammer, insert a Ficher strong anchor into the hole.
Ficher strong anchor
M8 x 30
Hammer
N/A
4. Using the hammer, insert an EHS tool into the inner bolt.
EHS tool
EAW H
8 x 30
Hammer
N/A