2. Choosing an Installation Site
Green 16/18 Installation Manual 15
En
glis
h
Minimum Space Required
Without CEPH unit
2,055 mm (L) x 2,495.8 mm (W) x 2,435 mm (H)
With CEPH unit
2,804 mm (L) x 2,495.8 mm (W) x 2,435 mm (H)
If the ceiling height is less than 2403.5 mm (without Base) /
2435.0 mm (with Base) (= max. height of the 100
mm), refer to
Appendix
C. Limiting the Column Height
to
lower the max. Height of the column.
Lead Thickness
≥
1 mm
Width of the Entrance
The door of the X-
r
ay room should have a clearance of more than 800 mm (31.5”) wide.
Floor Area
The floor of the X-
r
ay room must be stable and level for system balance.
The floor must be able to support a minimum weight of 500 kg/m
2
(110 lbs. / feet
2
)
Protection against Radiation
To protect against radiation hazards, follow all federal and municipal requirements.
During exposure, the operator should follow applicable radiation shielding
requirements and remain at least 2m (7’) from the source of the radiation.
Maintain visible contact with the patient and a clear view of indicators such as the
warning lamp and imaging status on the PC.