Calibration: Get Ready
The correct equipment is needed to calibrate Sprint LC. Accurate
calibration standards are necessary to proceed with calibration. Read the
complete specifications of the calibration standard intended for use to
ensure the standard is adequate for the task.
All calibration adjustments are handled through the firmware. No
potentiometers, dip switches, or other electrical components need to be
adjusted to calibrate the Sprint LC.
Exercise Caution
Once entering the calibration menus, the potential exists to negatively
influence tester accuracy. Calibration requires the right tools and
knowledge to do the job. Do not attempt a calibration until this section of
the guide is studied and the consequences of starting a calibration are
understood.
If Sprint LC is not due for a scheduled calibration, but it appears to have
an accuracy problem, check all other possibilities and consult with Uson
before starting a calibration. For example: - Verify or compare against a
standard first before changing calibration data. - Make sure the tester or
test fixtures do not leak.
Standards
Calibration standards must have the necessary range, accuracy, stability,
and repeatability to calibrate Sprint LC. Check the specifications of your
Sprint LC to make sure calibration standards have at least twice the
accuracy as that required by your particular Sprint LC.
Connections
To calibrate Sprint LC, air connections between Sprint LC and the
calibration standards must be leak-tight, direct, and as short as possible.
Leaks and flexing connections can cause calibration errors.