Certified Restrictor Used as Flow Master
FIGURE 83
A certified restrictor can be used to check the accuracy of the flow
transducer. A certified flow master calibrated to flow at multiple pressures
is called a flow linearizer. Obtain flow linearizers in the specific range of
the tester from Uson. These devices are shipped with a pressure-to-flow
chart and a certificate of accuracy.
Read the
pressure standard
when setting the pressure to produce the air
flow as specified in the pressure-to-flow chart. Always use the more
accurate standard to measure the pressure when using a flow linearizer.
If there is no access to a flow linearizer, use individual fixed restrictors.
These are often used when calibrating Sprint LC at a single flow point. A
single flow point calibration is sometimes acceptable when Sprint LC is
used at just one flow rate or for a simple verification.
Uson
SPR
INT L
C
1
2
Sprint LC
Flow Master
Pressure
Standard