![Uson SPRINT LC Owner'S Manual Download Page 46](http://html1.mh-extra.com/html/uson/sprint-lc/sprint-lc_owners-manual_1064343046.webp)
Page 46
Coupling Example
The Sprint LC tester can control an external fixture in which specific
actions must take place in a certain sequence before the leak or flow test is
started.
Coupling
In this example, three actions must take place BEFORE a test is started—
(1)
A shuttle (moving fixture) holding the product moves into position.
(2)
Air cylinders move down to clamp on top of the product.
(3)
Prex or FasTest seals connect air pressure to the bottom of the
product.
Uncoupling
AFTER the test is finished, Sprint LC uncouples in the reverse order set in
the coupling menu.
(1)
FasTest seals disconnect air pressure from the bottom of the
product.
(2)
Air cylinders move up to open the top of the product.
(3)
The shuttle holding the product moves back for reloading.
4-Way (Normally-Open) Valves
Sprint LC is generally supplied with three-way, normally closed solenoid
valves for coupling control. As a special option, Sprint LC can be ordered
with one, two, or three internal four-way coupling valves. In this case, the
coupling ports will be labeled to indicate which valve output is the
normally open connection and which is the normally closed connection. If
4-way valves are used, the normally closed side will work as described in
the coupling setup. The normally open valve connection will be a mirror
image and will operate in reverse of the normally closed valves.