Page 117
Fill time in a Flow Test
Use the fill time in a flow test to get the product pressurized
and flowing properly before taking flow readings. This
protects the flow sensor from overpressure and provides
more consistent flow readings.
FIGURE 57
Figure 57 shows the air flow measured
by the mass flow sensor is above the
maximum flow setpoint at the end of test
phase. Sprint LC turns on the reject light
and shows
HIGH
in the status box on the
run display.
If dump is active, the dump valve opens
and pressure is vented to atmosphere.
USER
TIP
.0 HIGH
Test Time Status
PROGRAM 1
Press. Tested Reject
93 psig 1 1
sccm
Fill
Test
Zero
Couple
High Flow Reject
TIME
FLOW
Min Flow
Max Flow
Flow is above Maximum setpoint
at the end of the test time.