Page 113
Flow Test Parameters
Parameters in a flow test—
Test Pressure
Pressure Error
Coupling Time (if external fixtures are used)
Fill Time (only required on large products)
Test Time
Dump (auto, off or time)
Max Flow (maximum flow)
Min Flow (minimum flow)
Next Program
Protect the Flow Sensor
Do NOT run a flow test unless product is attached to
the test port. Without this restriction, the high flow rate
could damage the flow sensor.
FIGURE 53
!
Fill
Test
Zero
Couple
Flow Test Setpoints
TIME
PR
ESS
U
R
E
& F
LOW
Min Flow
Max Flow
Test Pressure
(- P ERR)
(+ P ERR)
Flow and Pressure are compared to
Pass/Fail limits at the end of the test.