Page 141
Pressure Calibration
Setup
Sprint LC uses a lookup table to establish pressure transducer
linearization. It is necessary to set the zero point and multiple pressure
points along the range of the particular tester. The exact pressure
calibration points in the scale are entered at the factory into the calibration
menus.
When performing a pressure calibration, air is sent out of the Sprint LC to
supply a calibrated external standard for comparison to the Sprint LC.
First use the
INCREMENT
and
DECREMENT
keypads to select the target
pressure. Target increments are typically in ten or more steps, depending
on the range of the installed sensors.
Adjust the regulator on the back of the Sprint LC until the target number
very closely matches the external pressure standard.
When the external standard closely matches the target number, then store
the calibration data point.
Store Calibration Data Points
It is necessary to press and hold the START switch, and then press the
BACK keypad to store a new calibration setting. Both the START and the
oval BACK keypad must be active at the same time to store a new data
point.
Storing calibration data points operates the same way for pressure & flow
calibrations.
Valve Heat
Do not leave the tester in calibration mode. During
calibration the valves are held active. Allowing the
tester to remain in calibration mode too long will
allow the valves to create unwanted heat that can
adversely affect the calibration.
!