![Uson SPRINT LC Owner'S Manual Download Page 143](http://html1.mh-extra.com/html/uson/sprint-lc/sprint-lc_owners-manual_1064343143.webp)
Page 143
Pressure Calibration
Procedure
Access the flow calibration menu as shown in Figure 77 on page 140
1.
The zero is set first, so ensure the air supply going to Sprint LC is
disconnected and the pressure regulator knob is adjusted fully out
(off). Press the
DECREMENT
keypad so the target value on the
calibration menu reads 0.00.
2.
The reading value on the calibration menu should display a value
close to zero. The calibration standard connected to the Sprint LC
should also read zero. If not, make sure all air connections are
correct.
3.
Press and hold the white START switch while momentarily
pressing the
ENTER/BACK
keypad. The reading value on the
calibration menu should now read 0.00.
4.
Connect air pressure to the air input connector on the rear of the
tester. Press the
INCREMENT
keypad once to advance the target to
the next setpoint (5.00 psig in our example)
5.
Now, slowly adjust the pressure regulator. Watch the readout on
the
pressure standard
. (For the remainder of this procedure,
ignore the value called
reading
on the calibration menu unless
wanting to record the value as-found data). When the pressure
standard reads the target value, press and hold the white START
switch while momentarily pressing the
ENTER/BACK
keypad, to
accept this as a calibration number.
6.
Repeat step 5 for each target value programmed into the Sprint LC
until the maximum range is calibrated.
7.
The final step in the procedure is to verify that zero point has not
moved. Remove pressure from the air input, and turn the regulator
fully counterclockwise. If the zero reading is not within 25% of
tolerance then the calibration may need to be repeated.
Exit Calibration and allow the tester to sit for 15 minutes before
repeating the calibration. This will allow the valves to cool before