Page 34
Setup Menus
From the setup menu, it is possible get to four menus used to set operating
conditions. These four menus are called Test Parameters, Units,
Calibration, and Serial Port. A brief description of the setup menus
follows.
Test Parameters Menu
The test parameters menu is the heart of Sprint LC operations. This is
where is entered all information that tells Sprint LC how to decide whether
a product under test is accepted or rejected.
Options on the parameters menu change depending on whether you’re
setting up a pressure decay, occlusion, flow, or burst test. Please note that
not all test types are available in every tester.
Sprint LC comes with ninety nine stored programs—
f
Program 1
Pressure Decay
f
Program 2
Occlusion
f
Program 3
Flow
f
Program 4
Burst
f
Program 5
Pressure Decay
f
(Pressure decay continued for all 99 programs)
Factory-installed programs can be modified in any way. For instance, it is
possible to change all programs to flow measurement programs if desired.