Calibration
with
Additional Manual Regulator
Please refer to the calibration section of the
Sprint LC Owner’s Guide
.
This appendix describes only the differences when performing calibration
on testers using the additional manual regulator.
Two Channel Concurrent Sprint
The calibration functions much the same as with a single regulator. The
calibration for each test port (and sensor) is done independently in its own
calibration menu. Simply do the calibration as described for the first
channel and then repeat it for the second channel using the external
regulator.
Single Channel or Sequential Sprint
The calibration functions much the same as in a standard Sprint LC.
During a calibration the external or HIGH regulator is used to adjust the
calibration pressure. The internal regulator is not used in calibration. If
Sprint LC is within calibration tolerance during an accuracy check, Uson
recommends that no adjustment be made.
Setting test Pressure with 2
nd
Manual Regulator
Back Prev Next GoTo
PROGRAM 1
SET TEST PRESSURE
Regulator LOW
Test Press. 10.000 psig
Test Type PD
Test Press. GoTo
Press. Error .5 psig
Coupling GoTo
Test Params GoTo
Next Program 0
FIGURE 91
Appendix 10