- Specify the start point and end point of the section by displacement.
- After completion of measurement, the waveform is scanned from the front, and each hold
operation is performed by the set Section Start-End.
- Turning OFF
→
ON the external input signal “Start” releases the hold, and turns OFF all outputs
but Run. (Reset operation by the Start input)
- Turning OFF
→
ON the Stop signal or taking data of 2048 points turns ON the Complete signal.
- All results are output after completion of measurement.
- The OK signal is output when it is confirmed that judgment has been made at all sections and the
results are not HI or LO upon completion of measurement.
OK will not result if judgment is not made for the Number of Use Sections.
Example 1: Failure in hold detection when Relative Maximum, Relative Minimum, or
Inflection Point hold is used.
Example 2: Sections not changed for the Number of Use Sections.