UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:TIMebase:WINDow:OFFSet
Command format:
:TIMebase:WINDow:OFFSet <offset>
:TIMebase:WINDow:OFFSet?
Functional description:
For adjust WINDow offset value
(
zoom timebase <Zoomed>
)
,wave shift relative to screen center.
Return format:
Query return<offset> value by scientific notation method, unit is s.
For example:
:TIM:WIND:OFFS 1
set WINDow offset value as 1s
:TIM:WIND:OFFS?
query return 1.000e000
:TIMebase:SCALe
Command format:
:TIMebase:SCALe {<scale> | UP | DOWN}
:TIMebase:SCALe?
Functional description:
Set timebase position of MAIN, s/div.
<scale>
:
timebase value
;
UP
:
plus one position based on the current position
;
DOWN
:
minus one position based on the current position.
Return format:
Query return< scale> value by scientific notation method, unit is s/div.
For example:
:TIM:SCAL 2
set MAIN offset value as 2s/div
:TIM:SCAL?
query return 2.000e000
:TIMebase:WINDow:SCALe
Command format:
:TIMebase:WINDow:SCALe < scale >
:TIMebase:WINDow:SCALe?
Functional description:
For adjust WINDow timebase
(
zoom timebase <Zoomed>
)
, that is s/div.
Return format:
Query return < scale> value by scientific notation method, in s/div.
For example:
:TIM:WIND:SCAL 2
set WINDow timebase offset value as 2s/div
:TIM:WIND:SCAL?
query return 2.000e000