UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
Display format of grid, FULL
(
display grid and coordinate
)、
GRID
(
display grid
)、
CROSS
(
display coordinate
)、
NONE
(
not display grid and coordinate
)
.
Return format:
Query return{ FULL | GRID | CROSS | NONE}.
For example:
:DISPlay:GRID:MODE FULL
set display grid and coordinate
:DISPlay:GRID:MODE?
query return FULL
:DISPlay:CLEar
Command format:
:DISPlay:CLEar
Functional description:
Empty waveform display on the screen.
:DISPlay:TYPE
Command format:
:DISPlay:TYPE {XY12| XY34|YT}
:DISPlay:TYPE?
Functional description:
Set the display type of time base as XY12 (X-Y method: display amplitude of channel 1 on the
horizontal axis, display amplitude of channel 2 on the vertical axis). XY34 (X-Y method: display
amplitude of channel 3 on the horizontal axis, display amplitude of channel 4 on the vertical axis).
YT (Y-T method: display the relative relationship of vertical voltage and horiziontal time).
Return format:
Query return XY12, XY34 and YT.
For example:
:DISP:TYPE YT
set the time base as YT method
:DISP:TYPE?
query return YT
WAVeform Command
This command is to read waveform data and relevant parameters.
:WAVeform:FORMat
Command format:
:WAVeform:FORMat
{ WORD |
BYTE |
ASCII | CSV | DAT }
:WAVeform:FORMat?
Functional description: