UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
CHANnel Command
This command is set channel independently, <n> take value as 1/2/3/4/5/6/7/8/9, it
presents {CH1/CH2/ CH3/ CH4/ MATH/ REF-A/ REF-B/ REF-C/ REF-D}.
:CHANnel<n>:BWLimit
Command format:
:CHANnel<n>:BWLimit {{1|ON}|{0|OFF}}
:CHANnel<n>:BWLimit?
Functional description:
Set bandwidth limit function is ON
(
turn on bandwidth limit to 20MHz
,
to reduce display noise
)
or OFF
(
turn off bandwidth limit to achieve full display bandwidth
)
.
Return format:
Query return 1 or 0
,
it presents ON or OFF.
For example:
:CHAN1:BWL ON
turn on channel 1 bandwidth limit.
:CHAN1:BWL?
query return 1
,
it presents channel 1 bandwidth limit is turn on
:CHANnel<n>:COUPling
Command format:
:CHANnel<n>:COUPling {DC|AC|GND}
:CHANnel<n>:COUPling?
Functional description:
Set channel coupling mode, DC presents the AC and DC component of input signal; AC presents
blocked the DC component of input signal; GND presents cut off input signal.
Return format:
Query AC, DC or GND.
For example:
:CHAN1:COUP DC
set channel 1 coupling mode as DC
:CHAN1:COUP?
query return DC