Set I2C bus address, parameter is character string of 0x format, range is related with setup
parameter of command SBUS:I2C:ASIZe, which is [0~2^addressbit – 1].
Return format:
Query return character string address value of 0x format.
For example:
:SBUS:I2C:ADDRess "0x7F"
I2C bus address as 0x7F
:SBUS:I2C:ADDRess?
query return 0x7F
:SBUS:I2C:DATA:QUALifier
Command format:
:SBUS:I2C:DATA:QUALifier {EQUal | LESSthan | GREaterthan}
:SBUS:I2C:DATA:QUALifier?
Functional description:
Set data term of I2C bus.
Return format:
Query return{EQUal | LESSthan | GREaterthan}.
For example:
:SBUS:I2C:DATA:QUALifier EQUal
set I2C bus term as EQUal
:SBUS:I2C:DATA:QUALifier?
query return EQUal
:SBUS:I2C:DATA
Command format:
:SBUS:I2C:DATA <value>
:SBUS:I2C:DATA?
Functional description:
Set I2C bus data, parameter is character string of 0x format, data range is
0x0~0xFFFFFFFFFFFFFFFF.
Return format:
Query return data of character string of 0x format.
For example:
:SBUS:I2C:DATA "0xFFFFF"
set I2C bus data as 0xFFFFF
:SBUS:I2C:DATA?
query return 0xFFFFF
:SBUS:I2C:QUALifier
Command format:
:SBUS:I2C:QUALifier {STARt | RESTart | STOP | LOSS | ADDRess | DATA | ADATA}