Page 82 of 85 Pages
Message /
Symbol /
Problem
Failure Description
Corrective Action
pressure conditions in the chamber,
after preset time, during the air
removal stage.
RTC Error -
Please Set
Current Date
and Time
This message is displayed in order
to set the date and the time.
Set Current Date And Time.
If the problem persists, call the
technician.
Time Error
This message is displayed if one of
the real time clocks is faulty.
Call the technician.
Door is open
(During the
cycle)
This message is displayed when the
door is open: During the cycle.
Close the door to perform a
new cycle.
Canceled By
User
This message is displayed after the
START/STOP key is pressed and
cycle aborted.
Wait until "cycle failed
–
canceled by user" or "cycle
end
–
canceled by user" is
displayed.
Perform a new cycle.
Cycle Failed
This message and symbol are
displayed if an error occurs before
sterilization cycle is completed.
Perform a new cycle.
Air Error
This message is displayed at the end
of the cycle If the autoclave does
not reach the atmospheric pressure
after 10 minutes since the drying
stage started.
Wait until the autoclave
reaches the atmospheric
pressure and perform a new
cycle.
Compressed
air supply
error
(ELVCPVG-
D only)
This message is displayed in case of
a compressed air supply
malfunction.
Check and fix the compressed
air supply.
Periodical
check time
exceeded -
Please call for
service
The periodical maintenance time has
passed.
Call for service.
Cycle counter
exceeded -
Please call for
service
Number of cycles, since last
periodical maintenance, exceeded
the preset number as defined by
"cycle counter" parameter.
Call for service.
Power Down
This message is displayed if power
down has occurred during the cycle.
(this message will print out in the
printer after the autoclave will turn on).
Turn on the autoclave and wait
until the autoclave is ready
(reaches the safe condition)
and perform a new cycle.
Supply
distilled water
error (digitat
input option)
This message is displayed in case of
a mineral free water supply
malfunction.
Check and fix the mineral
free water supply