Program 13: Warm-Up
Pre-warming the chamber at 95
C (194
F) with no drying.
Caution
Validation of the sterility when using this program is the responsibility of
the user.
Nominal parameters default settings
Sterilization temperature: 95
C (194
F).
Sterilization time: 5 minutes.
Operations sequence:
Heating stage; steam enters the chamber and warms up until the warm
up temperature 95
C (194
F) is reached inside the chamber.
Warm up temperature is maintained constant for the preset sterilization
time.
Slow exhaust, steam is slowly exhausted from the chamber, until the
pressure equals the atmospheric pressure.