Program 6: Liquid B
–
Waste Cooling (ELC
–
cooling models only)
For liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC
.
for 30 minutes, such as liquid waste.
Nominal parameters default settings
Sterilization temperature: 121
C (150
F).
Sterilization time: 30 minutes.
Operations sequence:
(PV, pre-vacuum model only) Air removal stage; four vacuum pulses
removes air by drawing the chamber down to 20kPa.
Heating stage; steam enters the chamber and warms up until the
sterilization temperature is reached.
Sterilization temperature is maintained constant during the sterilization
time.
Forced cooling to the required end temperature (approx. 85ºC), while
the pressure in the chamber remains the same as the sterilization
pressure, by supplying compressed air into the chamber. The forced
cooling shortens the cooling stage significantly.
Fast exhaust, the steam is exhausted out of the chamber at a fast rate
until pressure decreases to ambient pressure.
Note: (BH, bio-hazard model only) If the cycle fails, fast exhaust is
performed through the bio-hazard filter.