Page 10 of 85 Pages
3
TECHNICAL DATA
3.1
Stand
–
by heating mode
The autoclave provides an option of heating the chamber in stand-by mode
between cycles with a very low power in order to reduce total cycle time
(1.6% of the total power only). The autoclave turns off automatically if the
interval between the sterilization cycles is more than 2 hours.
3.2
Storage Conditions
Packed or unpacked, the autoclave shall be retained in indoor
conditions!
3.3
Operating Conditions
The autoclave is intended to w
ork in ‘indoor’ conditions only.
Only autoclavable materials shall be used.
The environment shall not exceed an ambient temperature range of 5
º
C
(41
º
F)-40
º
C (104
º
F) and a relative humidity of 85% respectively.
The operation altitude shall not be over 2000 meters (6561 feet)
(ambient pressure shall not be lower than 80 kPa (11.6 psi)).
The autoclave shall not be used in a manner not specified in this
manual!
Do not use the autoclave in the presence of dangerous gases.
Operate the autoclave only in the manner specified in the manual. If
the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
Caution!
Waste water should be brought into the public net in accordance with
the local rules or requirements i.e ONLY NON-HAZARDOUS LIQUIDS
SHALL BE DISPOSED IN PUBLIC SEWAGE!
3.4
Utilities
3.4.1
Electrical utility
Power
Switch box
Recommended
Circuit Breaker
208-220V (3 Ph + Earthing)
No
20A
380-400V(3 Ph + N
Earthing)
No
16A
Frequency
50/60Hz