Page 22 of 85 Pages
4.2.3
Printer
The printer is an optional device.
It prints the detailed history of each cycle performed by the
autoclave. The printing is on thermal paper with 24 characters
per line and records the sterilization cycle information for
subsequent consideration.
4.3
Displayed Error Messages / Symbols
The failures are divided into two categories:
a.
Failure that occur before completing the sterilization stage, which in
this case will leave the load unsterilized
b.
Failure that occur after completing the sterilization stage, which in
this case will leave the load sterilized
For the list of
Displayed Error Messages / Symbols
see sec. 12. TROUBLESHOOTING
4.4
Displayed operational messages / Symbols
Message /
Symbol
Name
Message / Symbol
Description
Required Action
This symbol is displayed
when the door is open.
Close the door.
Door is open
(during stand
by)
This message is displayed
when the door is opened: In
standby - if START/STOP
is preset.
Close the door to perform a
new cycle.
Cycle Ended
This message is displayed
when the cycle ended
successfully.
Press START/STOP in
order to clear this message
and to allow door opening.
Test Ended
This message is displayed
when the test ended.
Press START/STOP in
order to clear this message
and to allow door opening.
This symbol is displayed
when Cycle by Clock mode
is performed.
Enter the Admin menu as
described in this manual to
change the time or to cancel
this option.
Cycle by
clock
This message is displayed if
the user presses
START/STOP key while
the "cycle by clock" mode
is active.
Enter the Admin menu as
described in this manual to
change the time or to cancel
this option.
Atmospheric
pressure not
set
This message is displayed in
order to set the atmosphere
pressure by opening the
door for 5 minutes.
Open the door for 5 minutes
in order to set the
Atmospheric pressure.