AK-Protocol
F-5
C o m m a n d D e t a i l s
Control Commands
Set Stand-by Mode
Command
Response
Description
STBY K0
STBY E
Set instrument to stand by-mode
(turn pumps and chargers OFF)
Set Measurement Mode
Command
Response
Description
SMEA K0
SMEA E
Set instrument to measurement
mode.
(turn pumps and chargers ON)
Write Commands
Write Particle Density Vector
Command
Response
Description
EPDV K0 B1 B2 B3…B32
EPDV E
Write particle size density to
instrument
BX = Particle density for bin X
Range: 0 to 10.00 [g/cm³.]
Format: Float
Write Column Heater Mode
Command
Response
Description
ECHM K0 ON
ECHM E
Turn column heater ON
ECHM K0 OFF
ECHM E
Turn column heater OFF
Write Data Simulation Mode
Command
Response
Description
EDSM K0 ON
EDSM E
Turn data simulation mode ON
EDSM K0 OFF
EDSM E
Turn data simulation mode OFF