D-8
Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
RF
Read Flow
RF
Read Flow (n text)
RF,#
Read Flow (numbers only)
RF,S
Read Flow Setpoint (n text)
RF,S,#
Read Flow Setpoint (numbers only)
RAV
Read Analyzer Voltage
RAV
Analyzer Voltage (n text)
RAV,#
Analyzer Voltage (numbers only)
RC
Read Charger
RC
Read Charger Current and Voltage (n
text)
Return: 'CHx' [0<=CHx<=1023]
RC,#
Read Charger Current and Voltage (numbers only)
RC,S
Read Charger Current and Voltage Setpoint
(n text)
RC,S,1,#
Read Negative Charger Current Setpoint
(numbers only)
RC,S,2,#
Read Positive Charger Current Setpoint (numbers
only)
RADC
Read Analog Digital Converter
RADC
Read all ADC Channels (n text)
RADC,#
Read all ADC Channels (numbers only)
RADC,x
Read ADC Channel x [0<=x<=23] (numbers only)
RT
Read Temperature
RT
Read Temperature (n text)
RT,#
Read Temperature (numbers only)
RIE
Read Instrument Error
RIE
Read Errors (n text)
RIE,#
Read Errors (numbers)
Bit Position
Error Condition
0
Sheath Flow Rate low
1
Sheath Flow Rate high
2
Sample Flow Rate low
3
Sample Flow Rate high
4
Charger Flow Rate low
5
Charger Flow Rate high
6
Extraction Flow Rate low
7
Extraction Flow Rate high
8
Sheath Flow Temperature low
9
Sheath Flow Temperature high
10
Sample Flow Temperature low
11
Sample Flow Temperature high
12
Charger Flow Temperature low