Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
Read Instrument Error
Command
Response
Description
AERR K0
AERR E S0 S1 S2…S63
Read current instrument error conditions. If no error
is detected the command returns
SX = 0:
No Error
SX = 1:
Error Detected
Format:
Text
Possible Error Conditions
S0 : Sheath Flow Rate low
S1 : Sheath Flow Rate high
S2 : Sample Flow Rate low
S3 : Sample Flow Rate high
S4 : Charger Flow Rate low
S5 : Charger Flow Rate high
S6 : Extraction Flow Rate low
S7 : Extraction Flow Rate high
S8 : Sheath Flow Temperature low
S9 : Sheath Flow Temperature high
S10 : Sample Flow Temperature low
S11 : Sample Flow Temperature high
S12 : Charger Flow Temperature low
S13 : Charger Flow Temperature high
S14 : Extraction Flow Temperature low
S15 : Extraction Flow Temperature high
S16 : Analyzer Voltage Upper Section low
S17 : Analyzer Voltage Upper Section high
S18 : Analyzer Voltage Middle Section low
S19 : Analyzer Voltage Middle Section high
S20 : Analyzer Voltage Lower Section low
S21 : Analyzer Voltage Lower Section high
S22 : Not used
S23 : Not used
S24 : Positive Charger Current low
S25 : Positive Charger Current high
S26 : Positive Charger Voltage low
S27 : Positive Charger Voltage high
S28 : Negative Charger Current low
S29 : Negative Charger Current high
S30 : Negative Charger Voltage low
S31 : Negative Charger Voltage high
S32 : Absolute Pressure low
S33 : Absolute Pressure high
S34 : Particle Concentration high
S35 : Not used
S36 : Not used
S37 : Instrument Matrix Readback Error
S38 : ELM Covariance Readback Error
S39 : ELM Offset Readback Error