Serial Commands
D-15
SELM
Set Electrometer Parameter
SELM,RESET
Reset all electrometers
Wait 30 seconds after a reset to take data
SELM,ZERO
Initiate countdown to zero the electrometer.
The countdown is updated every second
and takes 45 seconds to complete. On
completion the new offset and covariance
values are written to the DSP and an OK will
be returned
SELM,STOP
Stop the zeroing of electrometer
SAO
Set Analog Out
SAO,R,x
Set Voltage range [x=0(0-5) x=1(0-10)]
SAO,L,C,x
Set Lin/Log mode [x=0(Lin) x=1(Log)
SAO,N,C,x
Set Normalize mode [x=0(Off) x=1(On)
SAO,A,C,x
Set Averaging time [1<=x<=10]
SAO,D0,C,x
Set Minimum Diameter [5.624<=x<=562.3]
SAO,D1,C,x
Set Maximum Diameter [5.624<=x<=562.3]
Minimum Diameter must be < Maximum
Diameter
SAO,D,C,x
Set Density [x>=0.0]
SAO,M,C,x
Set Multiplier [x>=0.0]
SAO,E,C,x
Set Exponent [0.0<=x<=10.0]
SAO,F0,C,x
Set Minimum Full Scale [x>(=)0.0]
SAO,F1,C,x
Set Maximum Full Scale [x>0.0]
Min Full Scale must < Max Full Scale. Min
can be 0 in lin mode only
SAO,V,C,x
Set Voltage [0.0<=x<=5.0 or 10.0]
Sets Anout debug mode on. x=OFF sets
debug mode off
C = chan, 1-4
SDM
Set Debug Mode
SDM,x
Set Debug Mode [0<=x<=127]
SDM,0
Stop Dumping Data
SDM,2
Dump Sheath Flowrate [L/min]
SDM,3
Dump Sample Flowrate [L/min]
SDM,4
Dump Charger Flowrate [L/min]
SDM,5
Dump Extraction Flowrate [L/min]
SDM,6
Dump Sheath Flow Temperature [C]
SDM,7
Dump Sample Flow Temperature [C]
SDM,8
Dump Charger Flow Temperature [C]
SDM,9
Dump Extraction Flow Temperature [C]
SDM,10
Dump Sheath Flowrate [Bit]
SDM,11
Dump Sample Flowrate [Bit]
SDM,12
Dump Charger Flowrate [Bit]
SDM,13
Dump Extraction Flowrate [Bit]