5-38
Model 3090/3090AK Engine Exhaust Particle Sizer
TM
Spectrometer
Weighting symbols
. These are used to represent three predefined
values: Number(1), Surface(3.1416), and Volume(.5236).
Figure 5-31
Analog Out Setup
Voltage
Range
Select the voltage range for analog outputs. Either 0
–5 volts or
0-10 volts.
lin/log
Select whether the channel will display the analog voltage response
linearly proportional to concentration (lin) or with a logarithmic (Log)
response characteristic.
n
geSelectio
VoltageRan
MinConc
MaxConc
MinConc
nc
particleCo
V(lin)
geSelecion
VoltageRan
c)
Log(MinCon
c)
Log(MaxCon
)
LN(MinConc
eConc)
LN(particl
V(Log)
dN
dN/dlogDp
Select whether the channel will display the analog voltage response
“normalized” dN/dLog(Dp) or non-normalized (dN) over the
selected particle size range. dLog(Dp) = Log(MaxConc)
–
Log(MinConc).
Min/Max Dp
Spin Control but values can be entered manually. The Maximum
Value is 562.300, minimum is 5.624.
Power Dn