XD 7500
Operation
41
ba77191e01 - 71310060
04/2018
4.5.2
Measuring reagent tests, external barcode reader
For each method that can be measured with a reagent test, there is a cus-
tomary barcode. This is in the header line of the method description. In addi-
tion, it can also be downloaded from our website so that you can use it in your
own documents (e.g. a SOP).
Furthermore, on most reagent packagings, there is a barcode that indicates
the associated method. For reagents that can be used for several methods,
in this case the barcode indicates the most frequently-used method.
By scanning this barcode with the external barcode reader, the correspond-
ing method is selected (see also Section 3.4.5).
1
Open the cell shaft cover.
2
Scan barcode with the external
barcode reader. The photometer
selects the correct method with
the aid of the barcode.
3
Insert 24 mm round cell or rectan-
gular cell (see Section
4.2.5 I
). The cor-
rect measuring range is automati-
cally selected when the
rectangular cell (1, 2, 5 cm) is
inserted.
The photometer starts measuring
automatically.
<HOME>
Concentration
Concentration
16.01.17 9:52
Please select method for measuring
or insert a barcoded cell.
Setup
Method list
Letzte Methode
New Method