XD 7500
Operation
67
ba77191e01 - 71310060
04/2018
Variant 1:
Measure standard
solutions
1
Select and confirm
Measure stan-
dard solutions
.
2
Enter and confirm details of the
standard solutions (optional).
3
Use
[Next]
to apply all entries on
the page and move to the next
page.
The table for the measurement of
standard solutions pops up.
In the first two lines of the table,
the two value pairs (measuring
points) that are at least required
for a calibration are already pre-
pared (reagent blank value E0 and
any further nominal value).
4
Create further value pairs with
[Add]
as necessary.
You can delete a highlighted value
pair with
[Delete]
.
5
In the
Target value
column, enter
the nominal values of the individ-
ual standard solutions.
Measuring the standard solutions:
6
Using the arrow keys
<
><
>
and
<
><
>
, navigate to the rel-
evant input field in the
Absorbance
column and press
<START·ENTER>
.
Edit method
16.01.17 9:52
Standard ID
Standard manufacturer
Back
Next
Edit method
16.01.17 9:52
Target value
Absorbance
E0
0.000
1
Back
Add
Delete
Next
Edit method
16.01.17 9:52
Target value
Absorbance
E0
0.000
1
0.300
2
0.600
3
1.000
Back
Add
Delete
Next
Edit method
16.01.17 9:52
Target value
Absorbance
E0
0.000
1
0.300
2
0.600
3
1.000
Back
Add
Delete
Next