XD 7500
Technical data
163
ba77191e01 - 71310060
04/2018
Usable
Cells
Round cells 16 mm
– Outer diameter: 16 mm
– Inner diameter: 13.6 mm
– Flat cell bottom
– 13 mm with adapter
Round cells 24 mm
– Outer diameter: 24 mm
– Inner diameter: 21.5 mm
– Flat cell bottom
Round cells 13 mm
13 mm with adapter
Minimum filling level
20 mm
Minimum filling volume
Round cell 16 mm: 4 ml
Round cell 24 mm: 10 ml
Rectangular cell, 10 mm: 2 ml
Rectangular cell, 20 mm: 4 ml
Rectangular cell, 50 mm: 10 ml
Cell recognition
Automatic for most types
*
Depending on the wavelength range, different kinds of cells are suitable. Suitable are
round cells, all rectangular cells of glass, quartz or plastic, whose side surfaces are frosted
(see Section 8.1). Cells with clear or serrated lateral surfaces are not reliably recognized
by the automatic cell recognition.
Especially with plastic single-use cells we recommend you test them for suitability prior to
carrying out large-scale series of measurements.
For measurements in the UV range below 320 nm, glass cells and commercial PS plastic
cells are not suitable; below 280 nm, commercial PMMA plastic cells are not suitable due
to their transmission characteristics. Therefore, use quartz cells or tested single-use cells
(plastic) for applications in the UV range.
Measuring modes
Concentration
– Measurement with permanently programmed methods,
– Automatic method selection for test sets with barcode and external bar-
code reader
– Program support for the creation of additional user-defined methods
(max. 100)
– Citation forms and units method dependent
– Display of the absorbance value can be added
– Method data update possible via Internet