XD 7500
Operation
131
ba77191e01 - 71310060
04/2018
Locking of the photometer against all measurements (if activated).
Configuring the lock
of the photometer
Here you configure whether or not the photometer will be locked against all
measurements if there is no valid PCheck check or the interval for the
PCheck check has expired.
1
Enter a numeric value (2 to 52
weeks) (
<0...9>
) and confirm
The
Interval
defined for the
PCheck check is active.
<HOME>
DeviceCheck
–
PCheck setup
–
Interval
PCheck setup
16.01.17 9:52
Mode
Weeks
Interval
Lock instrument
Configuration of tests ...
Interval
2
Weeks
1
Select and confirm
Yes
.
The photometer is locked against
all measurements if the PCheck
check is invalid or the PCheck
interval has expired.
<HOME>
DeviceCheck
–
PCheck setup
–
Lock instrument
PCheck setup
16.01.17 9:52
Mode
Weeks
Interval
Lock instrument
Configuration of tests ...
Lock instrument
Should the instrument be locked for further
measurements if PCheck is invalid or has
expired?
No
Yes