Operation
XD 7500
134
ba77191e01 - 71310060
04/2018
Test record
A test record is displayed after the check. It can be printed and stored as a
file (in the internal DataB folder or USB memory device at the USB-A connec-
tion, see Section 4.11.1).
Sample printout:
5
Insert the cell.
The cell is automatically recog-
nized and the measurement
started.
After measuring, the measure-
ment result, Target value, Toler-
ance and an evaluation (OK or
failed) are displayed.
The photometer offers to repeat
the measurement if the check
failed.
With successful measurement,
the display shows the measure-
ment of the next test standard
from the Verification-Standard-Kit
XD, e.g. 430 LM.
6
Measure all test standards in the
same way.
After all test standards are suc-
cessfully measured, the check is
passed.
VERIFICATION STANDARD KIT XD 430 nm L 16.01.17 9:52
Please
VERIFICATION STANDARD KIT XD 430 nm L
XD 750009130512 2.70-Tintometer-0.14 Administrator
PCheck
OK
Protocol ID
9
Executed by:
Administrator
Executed
16.01.2017
Valid until:
16.02.2017
Verific.-Standard-Kit XD
OC479094
OK
430 L
0.205 +- 0.020
0.199
430 LM
0.402 +- 0.030
0.410
430 M
0.798 +- 0.040
0.801
430 H
1.610 +- 0.060
1.597
530 L
0.201 +- 0.020
0.203
.....
............
.....
.....
............
.....
(etc.)