Operation
XD 7500
132
ba77191e01 - 71310060
04/2018
Configuration of
tests ...
Carrying out the
PCheck.
(Example of Verifi-
cation Standard Kit
XD)
The PCheck includes the test with all test standards that were switched on
on the
DeviceCheck menu
/
PCheck setup
/
Configuration of tests ...
menu for
PCheck (see Page 132).
At the beginning there is a barcode test with the two test cells BCT1 and
BCT2 from the Verification Standard Kit XD. Then the test of the external bar-
code reader is done with a test barcode (included in the Verification Standard
Kit XD).
All configured test standards and
test standard sets are listed.
1
Select and confirm a test standard
or test standard set.
2
Adjust and confirm the extent of
the monitoring with
Switch on
or
Switch off
.
3
Confirm the test standard (set)
once again.
4
Switch to the adjustment of the
nominal values and tolerances
with
Setup
.
Example
Verification-Standard-Kit
XD
:
5
Using
<
><
>
and
<
><
>
,
select the
Lot number
,
Target
value
or
Tolerance
entries and
open them for editing with
<START·ENTER>
.
6
Enter and confirm the required
value (
<0...9>
)
7
Accept all values with
[Apply]
.
<HOME>
DeviceCheck
–
PCheck setup
–
Configuration of tests ...
PCheck setup
16.01.17 9:52
Verification Standard Kit XD
On
Secondary-Standard-Kit VIS Off
Secondary-Standard-Kit UV
Off
Inactive
Inactive
Inactive
I
Apply
Verification Standard Kit XD
General setup
Switch on
Verification Standard Kit XD
16.01.17 9:52
Lot number:
S0A1
Use by
16.04.2019
Target value
Tolerance
430 L
0.205
± 0.020
430 LM
0.402
± 0.030
430 M
0.798
± 0.040
430 H
1.610
± 0.060
530 L
0.201
± 0.020
530 LM
0.397
± 0.030
530 M
0.808
± 0.040
530 H
1.591
± 0.060
Apply
<HOME>
DeviceCheck
–
PCheck