Operation
XD 7500
142
ba77191e01 - 71310060
04/2018
9
Enter the volumes of sample and
standard of the individual test
sample solutions in the columns,
Sample [ml]
and
Standard [ml]
.
The nominal value is calculated
after each entry.
- Use
[Delete]
to remove a mea-
surement.
Note that all nominal values have
to be within the measuring range
of the test.
10
Use
[Next]
to apply all entries on
the page and move to the next
page. The entries are checked by
the photometer.
The photometer is ready to carry
out the measurements.
Carrying out measurements
According to the program, the
samples are measured top down.
You can, however, select the sam-
ples yourself and thus change the
order with
<
><
>
.
11
Use
[Measurement]
to proceed to
the measurement of the (first)
sample.
The measurement display is
shown.
12
Insert the cell with the respective
sample.
The sample is measured.
SCheck (Spike)
16.01.17 9:52
Method
130
Sample concentration
45 mg/lCSB
Sample
Standard
Target value
nominal
[ml]
[ml]
[mg/l]
[mg/l]
10
0.5
52.4
58
10
1
59.1
Back
Measureme
Complete
SCheck
16.01.17 9:52
Method
130
Sample concentration
45 mg/lCSB
Sample
10 ml
Standard
0.5 ml
To start measurement,
insert cell or press <START/ENTER>
16 mm
Back