Operation
XD 7500
68
ba77191e01 - 71310060
04/2018
The measurement display is
shown.
7
Insert the cell with the respective
standard.
The absorbance is measured. The
result of the first single measure-
ment is displayed.
8
If necessary, make additional sin-
gle measurements for the median
value formation with
[Next meas.]
or
discard the last single measure-
ment with
[Discard]
.
9
To accept the median value, press
[Apply]
.
If the zero standard concentration (reagent blank value E0) is not
measured and stored, the photometer calculates the calibration
line without this value. If the function
User-defined blank value
(in
the
Concentration
/
Setup
menu) is activated for measuring with
this method, the value for a0 is determined and replaces the cal-
culated axis intercept from the programming of the method (a0
see Page 64).
10
Repeat the steps 6 to 9 until all
input fields in the
Absorbance
col-
umn are filled out.
11
Use
[Next]
to apply all entries on
the page and move to the next
page.
The value pairs are displayed in a
diagram (standard: Polygon line).
Absorbance E0
16.01.17 9:52
To start measurement,
insert cell or press <START/ENTER>
525 nm
16 mm
Absorbance E0
16.01.17 9:52
Last measured absorbance
0.009
Median
0.009 (1 Measurement(s))
525 nm
16 mm
Next meas.
Discard
Apply
Edit method
16.01.17 9:52
Target value
Absorbance
E0
0.000
0.009
1
0.300
0.664
2
0.600
1.292
3
1.000
2.178
Back
Add
Delete
Next