XD 7500
Operation
65
ba77191e01 - 71310060
04/2018
Further method data
* mandatory inputs
** default value: mg/l
Input field
Possible entries
Number
*
1001 ... 1100
Designation
Arbitrary name (max. 18 characters)
Version
Any version designation (max. 18 characters)
Wavelength
*
Freely selectable (in nm)
Cell
*
13, 16, 24 mm (rund), 10, 20 or 50 mm
Citation form
e.g. PO4-P (max. 18 characters)
Unit
**
e.g. mg/l (max. 18 characters)
Resolution
*
0.001, 0.01, 0.1 or 1
Lower and upper limit of
the measuring range *
Any value between zero and the highest concen-
tration of the used standard solutions
Timer 0 to 3
Up to four analysis timers freely adjustable
MCheck target value
Any value within the measuring range
MCheck tolerance
Any
Required measure-
ments
1 or greater
Number of measurements after which a measured
value is documented. With more than one mea-
surement, the documented value is the median
from all measurements.
Reagent blank value
required
Yes
/
No
User calibration possible Yes
/
No
User calibration required Yes
/
No
If a nonlinear calibration line is programmed for a method, it may
be the case that the default setting of the following menu items
cannot be changed:
Reagent blank value required
User calibration possible
User calibration required