Operation
XD 7500
138
ba77191e01 - 71310060
04/2018
Test record
A test record is displayed after the check. It can be printed and stored as a
file (in the internal DataB folder or USB memory device at the USB-A connec-
tion, see Section 4.11.1).
Sample printout:
1
Via the DeviceCheck menu, select
the MCheck test and then the
method to be tested.
2
Carry out the check like a normal
measurement (see section 4.5.1
to 4.5.2).
3
Insert a cell or
Start the measurement with
<START·ENTER>
.
After the measurement is com-
pleted, the result and its evalua-
tion are displayed.
If the check failed, it is possible to
repeat the measurement.
If the check was successful, the
MCheck
function is finished.
MCheck check
16.01.17 9:52
[ZERO 15.01.17 11:11]
Target value
2.00 + 0.20 mg/l
To start measurement,
insert cell or press <START/ENTER>
65: Ammonium LR TT
NH
4
-N
16 mm
0.02 - 2.50 mg/l
Citation form
Unit
Later you can view the last test records for all MCheckmethods
monitored with MCheck under
MCheck info
.
09130512 2.70-Tintometer-0.14 Administrator
MCheck
OK
Protocol ID
32
Executed by:
Administrator
Executed
16.01.2017
Valid until:
13.03.2017
Method
65 NH4-N
Standard ID
VC 48201425
Target value
2.00 +- 0.20 mg/l
Measured value
2.14 mg/l